DiaSys respons®A1c
The respons®A1c is a specialized analyzer designed for the efficient measurement of HbA1c. Utilizing boronate affinity chromatography, the system provides precise results from minimal sample volumes, making it suitable for professional laboratory and point-of-care environments.
| Feature | Specification |
| Test Method | Boronate affinity chromatography |
| Sample Volume | 3.5 µL |
| Sample Material | Capillary or anticoagulated venous whole blood |
| Test Time | 4 minutes 30 seconds |
| Measuring Range | 4.0 – 15.0% DCCT (9 – 140 mmol/mol IFCC) |
| Hematocrit Range | 25 – 60% |
| Traceability | IFCC and NGSP |
| Interface | 4.3” colored touch screen with voice guidance |
| Memory | 400 results |
| Connectivity | 1 x USB-A, 1 x USB-B, 1 x Ethernet |
| Dimensions | 205 (W) x 200 (L) x 150 (H) mm |
| Weight | 1.6 kg |
| Item | Catalog Number | Size |
| respons®A1c Analyzer | 963100 | 1 Unit |
| respons®A1c Hemoglobin A1c Test Kit | 963207 | 10 tests |
Contact Björn Fetlaar your Diagnostics expert
Resources
Have product questions?
Brand

Simple 3-Step Testing
Streamlined Workflow
Step 1: Initialize
Insert the pre-calibrated cassette into the analyzer's cassette holder to prepare the system for measurement.
Step 2: Collect
Use the integrated capillary of the reagent pack to collect a precise 3.5 µL sample from a finger prick or venous blood.
Step 3: Analyze
Insert the reagent pack into the analyzer and close the lid; the system automatically delivers results in 4.5 minutes.
- Results in only 4.5 minutes
- Boronate affinity chromatography
- Room temperature reagent storage
- NGSP and IFCC traceable
- Results in only 4.5 minutes
- Boronate affinity chromatography
- Room temperature reagent storage
- NGSP and IFCC traceable
Frequently Asked Questions about respons®A1c Performance
Detailed technical information regarding the performance characteristics and handling of the respons®A1c HbA1c test.
The system utilizes boronate affinity chromatography for the determination of HbA1c.
The analyzer requires a minimal sample volume of 3.5 µL and provides reliable results within 4 minutes and 30 seconds.
The measuring range is 4.0 – 15.0% DCCT, which corresponds to 9 – 140 mmol/mol IFCC.
The test kits have a shelf life of 18 months from the date of production when stored under controlled conditions between 2 and 30°C.
The system demonstrates high precision (within series) with a CV of 2.6% at 5.4% DCCT and 1.9% at 9.0% DCCT, validated over 20 days using two analyzers and three reagent lots.